372 Chapter5
Language Reference
TRIGger Subsystem
Front Panel External Trigger Level
:TRIGger[:SEQuence]:EXTernal[1]:LEVel <voltage>
:TRIGger[:SEQuence]:EXTernal[1]:LEVel?
Set the trigger level when using the front panel external trigger input.
Factory Preset
and *RST: 2.0 V
Range: 5.0 to +5.0 V
Default Unit: volts
Front Panel
Access:
Mode Setup, Trigger, Ext Front, Level
Front Panel External Trigger Slope
:TRIGger[:SEQuence]:EXTernal[1]:SLOPe NEGative|POSitive
:TRIGger[:SEQuence]:EXTernal[1]:SLOPe?
Sets the triggering to occur on a positive-going edge or a negative-going
edge of the trigger when using the front panel external trigger input.
Factory Preset
and *RST: Positive
Front Panel
Access:
Mode Setup, Trigger, Ext Front, Slope
Rear Panel External Trigger Delay
:TRIGger[:SEQuence]:EXTernal2:DELay <time>
:TRIGger[:SEQuence]:EXTernal2:DELay?
Set the trigger delay when using the rear panel external trigger.
Factory Preset
and *RST: 0.0 s
Range: 500.0 ms to 500.0 ms
100.0 ms to 500.0 ms for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit: seconds
Front Panel
Access:
Mode Setup, Trigger, Ext Rear, Delay

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals