Chapter 5 309
Language Reference
SENSe Subsystem
RF Port Power Range Maximum Total Power
[:SENSe]:POWer[:RF]:RANge[:UPPer] <power>
[:SENSe]:POWer[:RF]:RANge[:UPPer]?
Set the maximum expected total power level at the radio unit under
test. This value is ignored if RF port power range is set to auto.
External attenuation required above 30 dBm.
Factory Preset
and *RST: 15.0 dBm
Range: 100.0 to 80.0 dBm for EDGE, GSM
100.0 to 27.7 dBm for cdmaOne, iDEN
200.0 to 50.0 dBm for NADC, PDC
200.0 to 100.0 dBm for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & Arib)
Default Unit: dBm
Remarks: Global to the current mode. This is coupled to the RF
input attenuation
You must be in the Service, cdmaOne,
EDGE(w/GSM),GSM, NADC, PDC, cdma2000,
W-CDMA (3GPP), or W-CDMA (Trial & Arib) mode to
use this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
Input, Max Total Pwr (at UUT)

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals