350 Chapter5
Language Reference
STATus Subsystem
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
register is cleared.
Operation Negative Transition
BASE
:STATus:OPERation:NTRansition?
This command determines what bits in the Operation Condition
register will set the corresponding bit in the Operation Event register
when the condition register bit has a negative transition (1 to 0). The
variable <number> is the sum of the decimal values of the bits that you
want to enable.
Factory Preset
and *RST: 0
Range: 0 to 32767
Operation Positive Transition
BASE
:STATus:OPERation:PTRansition?
This command determines what bits in the Operation Condition
register will set the corresponding bit in the Operation Event register
when the condition register bit has a positive transition (0 to 1). The
variable <number> is the sum of the decimal values of the bits that you
want to enable.
Factory Preset
and *RST:
Range: 0 to 32767

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals