EasyManuals Logo

Agilent Technologies E4406A VSA Series User Manual

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #320 background imageLoading...
Page #320 background image
320 Chapter5
Language Reference
SENSe Subsystem
Radio Format (Standard)
[:SENSe]:RADio:FORMat ARIB|TGPP|TRIal
[:SENSe]:RADio:FORMat?
Select the format that testing will be compliant with when
measurements are made.
ARIB, is the standard format defined by the Association of Radio
Industries and Business in Japan
TGPP, is the standard format defined by the Third Generation
Partnership Projects (3GPP)
Trial, is a 1998 trial format being evaluated
Factory Preset
and *RST: Trial
Remarks: You must be in the W-CDMA (Trial & Arib) mode to use
this command. Use INSTrument:SELect to set the
mode.
History: Version A.03.00 or later
Front Panel
Access:
Mode Setup, Radio, Standard
Radio Format (Standard)
[:SENSe]:RADio:FORMat M16QAM|M64QAM|DJSMR
[:SENSe]:RADio:FORMat?
Select the format that testing will be compliant with when
measurements are made.
M16QAM, is the standard iDEN format defined by Motorola
M64QAM, is the standard iDEN format defined by Motorola
DJSMR, is Japanese standard format that is based on the ARIB
RCR-32A standard
Factory Preset
and *RST: M16QAM
Remarks: You must be in the iDEN mode to use this command.
Use INSTrument:SELect to set the mode.
History: Version A.03.00 or later
Front Panel
Access:
Mode Setup, Radio, Format

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals