376 Chapter5
Language Reference
TRIGger Subsystem
Video (IF) Trigger Delay
:TRIGger[:SEQuence]:IF:DELay <time>
:TRIGger[:SEQuence]:IF:DELay?
Set the trigger delay when using the IF (video) trigger (after the
Resolution BW filter).
Factory Preset
and *RST: 0.0 s
Range: 500.0 ms to 500.0 ms
100.0 ms to 500.0 ms for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit: seconds
Front Panel
Access:
Mode Setup, Trigger, Video (IF Envlp), Delay
Video (IF) Trigger Level
:TRIGger[:SEQuence]:IF:LEVel <power>
:TRIGger[:SEQuence]:IF:LEVel?
Set the trigger level when using the IF (video) trigger.
Factory Preset
and *RST: 6.0 dBm for cdmaOne, GSM, Basic, Service,
cdma2000, W-CDMA (3GPP), W-CDMA (Trial & ARIB)
20.0 dBm for iDEN
30.0 dBm for NADC, PDC
Range: 200.0 to 50.0 dBm
Default Unit: dBm
Front Panel
Access:
Mode Setup, Trigger, Video (IF Envlp), Level