Chapter 5 327
Language Reference
SENSe Subsystem
Spectrum—Average Clear
[:SENSe]:SPECtrum:AVERage:CLEAr
The average data is cleared and the average counter is reset.
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Number of Averages
[:SENSe]:SPECtrum:AVERage:COUNt <integer>
[:SENSe]:SPECtrum:AVERage:COUNt?
Set the number of ‘sweeps’ that will be averaged. After the specified
number of ‘sweeps’ (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST: 25
Range: 1 to 10,000
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Averaging State
[:SENSe]:SPECtrum:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:SPECtrum:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST: On
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.