Page #378 background image
378 Chapter5
Language Reference
TRIGger Subsystem
RF Burst Trigger Level
:TRIGger[:SEQuence]:RFBurst:LEVel <percent>
:TRIGger[:SEQuence]:RFBurst:LEVel?
Set the trigger level when using the RF Burst (wideband) Trigger. The
value is relative to the peak of the signal. RF Burst is also known as RF
Envelope.
Factory Preset
and *RST: 6.0 dB
Range: 25.0 to 0.0 dB
200.0 to 0.0 dB for NADC, PDC
Default Unit: dB
Front Panel
Access:
Mode Setup, Trigger, RF Burst, Peak Level
RF Burst Trigger Slope
:TRIGger[:SEQuence]:RFBurst:SLOPe NEGative|POSitive
:TRIGger[:SEQuence]:RFBurst:SLOPe?
Set the trigger slope when using the RF Burst (wideband) Trigger.
Factory Preset
and *RST: Positive
Remarks: You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), or W-CDMA (Trial & ARIB) mode to use this
command. Use :INSTrument:SELect to set the mode.
Front Panel
Access:
Mode Setup, Trigger, RF Burst, Slope

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals