338 Chapter5
Language Reference
SENSe Subsystem
Burst Sync Delay
[:SENSe]:SYNC:BURSt:DELay <time>
[:SENSe]:SYNC:BURSt:DELay?
Set the delay for the burst measurement position from the reference
position that is determined by sync word or the burst rising/falling
edges.
Factory Preset
and *RST: 0 sec
Range: 500 ms to 500 ms
Default Unit: seconds
Remarks: You must be in the iDEN, NADC or PDC mode to use
this command. Use INSTrument:SELect to set the
mode.
Sync Burst RF Amplitude Delay
[:SENSe]:SYNC:BURSt:RFAMplitude:DELay <time>
[:SENSe]:SYNC:BURSt:RFAMplitude:DELay?
Set the delay for the RF amplitude sync.
Factory Preset
and *RST: 0 s
Range: 100 ms to 100 ms
Default Unit: seconds
Remarks: Global to the current mode.
You must be in the EDGE(w/GSM), GSM mode to use
this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
Mode Setup, Trigger, RF Burst, Delay

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals