EasyManuals Logo

Agilent Technologies E4406A VSA Series User Manual

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #207 background imageLoading...
Page #207 background image
Chapter 5 207
Language Reference
CALibration Subsystem
Internal 50 MHz Amplitude Reference Alignment Control
:CALibration:REF50[:DOIT]
:CALibration:REF50[:DOIT]?
Does automatic alignment of the internal 50 MHz amplitude reference
oscillator. You do this by setting an external source to 25.00 dBm and
using a power meter to measure the exact value. Then use
CAL:REF50:AMPL to input the source amplitude, measured on the
power meter. Finally, connect the source to the instrument RF INPUT
port and run the adjustment.
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command
Front Panel
Access:
System, Alignments, Align subsystem, Align 50 MHz
Reference
Enter Interactive Mode for Internal 50 MHz Amplitude
Reference Alignment
:CALibration:REF50:ENTer
Turns on the interactive mode for alignment of the internal 50 MHz
amplitude reference signal. Use CAL:REF50:ANOW to do the
alignment and CAL:REF50:EXIT to exit the interactive mode.
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command.
Front Panel
Access:
System, Alignments, Align subsystem, Align 50 MHz
Reference

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals