314 Chapter5
Language Reference
SENSe Subsystem
Power vs. Time—Resolution BW
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution]?
Set the resolution BW. This is an advanced control that normally does
not need to be changed. Setting this to a value other than the factory
default, may cause invalid measurement results.
Factory Preset
and *RST: 500 kHz
Range: 1 kHz to 5 MHz
Default Unit: Hz
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—RBW Filter Type
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution]:TYPE
FLATtop|GAUSsian
[:SENSe]:PVTime:BANDwidth|BWIDth[:RESolution]:TYPE?
Select the type of resolution BW filter. This is an advanced control that
normally does not need to be changed. Setting this to a value other than
the factory default, may cause invalid measurement results.
Flattop - a filter with a flat amplitude response, which provides the
best amplitude accuracy.
Gaussian - a filter with Gaussian characteristics, which provides the
best pulse response.
Factory Preset
and *RST: Gaussian
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.