212 Chapter5
Language Reference
DISPlay Subsystem
Display Annotation Title On/Off
:DISPlay:ANNotation:TITLe[:STATe] OFF|ON|0|1
:DISPlay:ANNotation:TITLe[:STATe]?
Turns the display of the title annotation on or off.
Front Panel
Access:
Display, Title, Title On Off
Turn the Entire Display On/Off
:DISPlay:ENABle OFF|ON|0|1
:DISPlay:ENABle?
Controls the updating of the display. If enable is set to off, the display
will appear to “freeze” in its current state. Measurements may run
faster if the instrument doesn’t update the display after every data
acquisition. There is often no need to update the display information
when using remote operation.
Factory Preset
and *RST: On
Remarks: The following key presses will turn display enable back
on:
1. If in local, press any key
2. If in remote, press the local (system) key
3. If in local lockout, no key
Front Panel
Access:
none
Error Vector Magnitude - View Selection
:DISPlay:EVM:VIEW POLar|CONStln|QUAD
:DISPlay:EVM:VIEW?
Select the view of EVM measurement
Factory Preset
and *RST: POLar
Remarks: You must be in the NADC or PDC mode to use this
command. Use INSTrument:SELect to set the mode.