246 Chapter5
Language Reference
MEASure Group of Commands
Power Statistics CCDF Measurement
This is a statistical power measurement of the complimentary
cumulative distribution function (CCDF). You must be in the Basic,
cdma2000, W-CDMA (3GPP), or W-CDMA (Trial & Arib) mode to use
these commands. Use INSTrument:SELect to set the mode.
The general functionality of CONFigure, FETCh, MEASure, and READ
are described at the beginning of this section. See the SENSe:PSTat
commands for more measurement related commands.
:CONFigure:PSTatistic
:FETCh:PSTatistic[n]?
:READ:PSTatastic[n]?
:MEASure:PSTatastic[n]?
History: Version A.03.00 or later, added in Basic A.04.00
Front Panel
Access:
Measure, Power Stat CCDF
After the measurement is selected, press Restore Meas
Defaults to restore factory defaults.
Measurement Results Available
n
0 Returns unprocessed I/Q trace data, as a series of comma-separated trace
points, in volts. The I values are listed first in each pair, using the 0 through
even-indexed values. The Q values are the odd-indexed values,
not specified or
n=1
Returns 10 comma-separated scalar results:
1. Average input power (in dBm)
2. Probability at the average input power level (in %)
3. Power level that has 10% of the power
4. Power level that has 1% of the power
5. Power level that has 0.1% of the power
6. Power level that has 0.01% of the power
7. Power level that has 0.001% of the power
8. Power level that has 0.0001% of the power
9. Peak power (in dB)
10.Count