Chapter 5 307
Language Reference
SENSe Subsystem
Center Frequency Step Size
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Specifies the center frequency step size.
Factory Preset
and *RST: 5.0 MHz
1.25 MHz for cdma2000
Range: 1.0 kHz to 1.0 GHz, in 10 kHz steps
Default Unit: Hz
History: Version A.03.00 or later
Front Panel
Access:
FREQUENCY/Channel, CF Stepl
RF Port Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_power>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the RF input attenuator. This value is set at its auto value if input
attenuation is set to auto.
Factory Preset
and *RST: 0 dB
12 dB for iDEN
Range: 0 to 40 dB
Default Unit: dB
Front Panel
Access:
Input, Input Atten