Chapter 5 345
Language Reference
SENSe Subsystem
Waveform—Control Decimation of Waveform Display
[:SENSe]:WAVeform:DECimate:STATe OFF|ON|0|1
[:SENSe]:WAVeform:DECimate:STATe?
Set the amount of data decimation done by the hardware in order to
decrease the number of acquired points in a long capture time. This is
the amount of data that the measurement ignores.
Factory Preset
and *RST: Off
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Sweep (Acquisition) Time
[:SENSe]:WAVeform:SWEep:TIME <time>
[:SENSe]:WAVeform:SWEep:TIME?
Set the measurement acquisition time. It is used to specify the length of
the time capture record.
Factory Preset
and *RST: 2.0 ms
10.0 ms, for NADC, PDC
15.0 ms, for iDEN mode
Range: 1 µs to 100 s
Default Unit: seconds
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.