Chapter 5 345
Language Reference
SENSe Subsystem
Waveform—Control Decimation of Waveform Display
[:SENSe]:WAVeform:DECimate:STATe OFF|ON|0|1
[:SENSe]:WAVeform:DECimate:STATe?
Set the amount of data decimation done by the hardware in order to
decrease the number of acquired points in a long capture time. This is
the amount of data that the measurement ignores.
Factory Preset
and *RST: Off
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Sweep (Acquisition) Time
[:SENSe]:WAVeform:SWEep:TIME <time>
[:SENSe]:WAVeform:SWEep:TIME?
Set the measurement acquisition time. It is used to specify the length of
the time capture record.
Factory Preset
and *RST: 2.0 ms
10.0 ms, for NADC, PDC
15.0 ms, for iDEN mode
Range: 1 µs to 100 s
Default Unit: seconds
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals