250 Chapter5
Language Reference
MEASure Group of Commands
Sensor Measurement
This checks the output of three sensors in the RF and IF circuitry. You
must be in the Service mode to use these commands. Use
INSTrument:SELect to set the mode.
The general functionality of CONFigure, FETCh, MEASure, and READ
are described at the beginning of this section.
:CONFigure:SENSors
:FETCh:SENSors[n]?
:READ:SENSors[n]?
:MEASure:SENSors[n]?
Front Panel
Access: With Service Mode selected,
Measure, Sensors
Measurement Results Available
n Results Returned
0 Not valid
not specified or n=1 Returns the following comma-separated scalar results:
1. IF signal amplitude is the ADC value for the detected 21.4 MHz IF
signal at the input to the analog IF.
2. Calibration Oscillator Level is a floating point number (is not
implemented, currently returns a zero).
3. RF temperature is a floating point number for the current
temperature in the RF section (in degrees Celsius).

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals