EasyManuals Logo

Agilent Technologies E4406A VSA Series User Manual

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #323 background imageLoading...
Page #323 background image
Chapter 5 323
Language Reference
SENSe Subsystem
Reference Oscillator External Frequency
[:SENSe]:ROSCillator:EXTernal:FREQuency <frequency>
[:SENSe]:ROSCillator:EXTernal:FREQuency?
Set to the frequency of the external reference oscillator being supplied
to the instrument. Switch to the external reference with ROSC:SOUR.
Preset
and *RST: Value remains at last user selected value (persistent)
Factory default, 10 MHz
Range: 1 MHz to 40 MHz, with 1 Hz steps
Default Unit: Hz
Remarks: Global to system
Front Panel
Access:
System, Reference, Ref Oscillator
Reference Oscillator Rear Panel Output
[:SENSe]:ROSCillator:OUTPut[:STATe] OFF|ON|0|1
[:SENSe]:ROSCillator:OUTPut?
Turn on and off the 10 MHz frequency reference signal going to the rear
panel.
Preset
and *RST: Persistent State with factory default of On
Remarks: Global to system. Was SENS:ROSC:REAR
Front Panel
Access:
System, Reference, 10 MHz Out

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals