262 Chapter5
Language Reference
SENSe Subsystem
SENSe Subsystem
Sets the instrument state parameters so that you can measure the
input signal.
Adjacent Channel Power Measurement
Commands for querying the adjacent channel power measurement
results and for setting to the default values are found in the “MEASure
Group of Commands” on page 233. The equivalent front panel keys for
the parameters described in the following commands, are found under
the
Meas Setup key, after the ACP or ACPR measurement has been
selected from the
MEASURE key menu.
Adjacent Channel Power—Average Count
[:SENSe]:ACP:AVERage:COUNt <integer>
[:SENSe]:ACP:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the average mode (termination
control) setting determines the average action.
Factory Preset
and *RST: 10 for cdma2000, W-CDMA (3GPP), W-CDMA (Trial &
Arib)
20 for Basic, cdmaOne, iDEN
Range: 1 to 10,000
Remarks: Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Averaging State
[:SENSe]:ACP:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACP:AVERage[:STATe]?
Turn average on or off.
Factory Preset
and *RST: On
Off for iDEN mode
Remarks: Use INSTrument:SELect to set the mode.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals