328 Chapter5
Language Reference
SENSe Subsystem
Spectrum—Averaging Mode
[:SENSe]:SPECtrum:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:SPECtrum:AVERage:TCONtrol?
Select the type of termination control used for the averaging function.
This determines the averaging action after the specified number of
‘sweeps’ (average count) is reached.
Exponential - Each successive data acquisition after the average
count is reached, is exponentially weighted and combined with the
existing average.
Repeat - After reaching the average count, the averaging is reset and
a new average is started.
Factory Preset
and *RST: Exponential
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Averaging Type
[:SENSe]:SPECtrum:AVERage:TYPE
LOG|MAXimum|MINimum|RMS|SCALar
[:SENSe]:SPECtrum:AVERage:TYPE?
Select the type of averaging.
Log The log of the power is averaged. (This is also known as video
averaging.)
Maximum The maximum values are retained.
Minimum The minimum values are retained.
RMS The power is averaged, providing the rms of the voltage.
Scalar The voltage is averaged.
Factory Preset
and *RST: Log
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals