Chapter 5 375
Language Reference
TRIGger Subsystem
Frame Trigger Synchronization Offset
:TRIGger[:SEQuence]:FRAMe:SYNCmode:OFFSet <time>
Lets you adjust the frame triggering with respect to the external trigger
input that you are using.
Factory Preset
and *RST: 0.0 s
Range: 0.0 to 10.0 s
Default Unit: seconds
History: Revision A.03.27 or later
Front Panel
Access:
Mode Setup, Trigger, Frame Timer, Offset
Trigger Holdoff
:TRIGger[:SEQuence]:HOLDoff <time>
:TRIGger[:SEQuence]:HOLDoff?
Set the holdoff time between triggers. After a trigger, another trigger
will not be allowed until the holdoff time expires. This parameter
affects all trigger sources.
Factory Preset
and *RST: 0.0 s
20.0 ms for iDEN
10.0 ms for NADC or PDC
Range: 0.0 to 500.0 ms
Default Unit: seconds
Front Panel
Access:
Mode Setup, Trigger, Trig Holdoff