Chapter 5 315
Language Reference
SENSe Subsystem
Power vs. Time—Sweep Time
[:SENSe]:PVTime:SWEep:TIME <integer>
[:SENSe]:PVTime:SWEep:TIME?
Set the number of slots which are used in each data acquisition. Each
slot is approximately equal to 570 ms. The measurement is made for a
small additional amount of time (about 130 µs) in order to view the
burst edges.
Factory Preset
and *RST: 1
Range: 1 to 50 (for resolution BW = 500 kHz)
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—Trigger Source
[:SENSe]:PVTime:TRIGger:SOURce EXTernal[1]|EXTernal2
|FRAMe|IF|IMMediate|RFBurst
[:SENSe]:PVTime:TRIGger:SOURce?
Select the trigger source used to control the data acquisitions.
External 1 - front panel external trigger input
External 2 - rear panel external trigger input
Frame - uses the internal frame timer, which has been synchronized
to the selected burst sync.
IF - internal IF envelope (video) trigger
Immediate - the next data acquisition is immediately taken,
capturing the signal asynchronously (also called Free Run).
RF Burst - wideband RF burst envelope trigger that has automatic
level control for periodic burst signals.
Factory Preset
and *RST: RF burst
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.