304 Chapter5
Language Reference
SENSe Subsystem
Correction for Base Station RF Port External
Attenuation
[:SENSe]:CORRection:BS[:RF]:LOSS <rel_power>
[:SENSe]:CORRection:BS[:RF]:LOSS?
Set the correction equal to the external attenuation used when
measuring base stations.
Factory Preset
and *RST: 0 dB
Range: 0 to 100 dB for cdmaOne
50 to 50 dB for Basic, iDEN, NADC or PDC
Default Unit: dB
Remarks: You must be in the Basic, iDEN, cdmaOne, NADC or
PDC mode to use this command. Use
INSTrument:SELect to set the mode.
Value is global to the current mode.
Correction for BTS RF Port External Attenuation
[:SENSe]:CORRection:BTS[:RF]:LOSS <rel_power>
[:SENSe]:CORRection:BTS[:RF]:LOSS?
Set equal to the external attenuation used when measuring base
transmit stations.
Factory Preset
and *RST: 0.0 dB
Range: 0.0 to 100.0 dB for GSM
50.0 to 50.0 dB for cdma2000, W-CDMA (3GPP), or
W-CDMA (Trial & Arib)
Default Unit: dB
Remarks: Global to the current mode.
You must be in the EDGE(w/GSM), GSM,
cdma2000,W-CDMA (3GPP), or W-CDMA(Trial & Arib)
mode to use this command. Use INSTrument:SELect to
set the mode.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals