Chapter 5 371
Language Reference
TRIGger Subsystem
Automatic Trigger Time
:TRIGger[:SEQuence]:AUTO[:TIME] <time>
:TRIGger[:SEQuence]:AUTO[:TIME]?
After the measurement is activated the instrument will take a data
acquisition immediately upon receiving a signal from the selected
trigger source. If no trigger signal is received by the end of the time
specified in this command, a data acquisition is taken anyway.
TRIGger[:SEQuence]:AUTO:STATE must be on.
Factory Preset
and *RST: 100.0 ms
Range: 1.0 ms to 1000.0 s
0.0 to 1000.0 s for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit: seconds
Front Panel External Trigger Delay Value
:TRIGger[:SEQuence]:EXTernal[1]:DELay <time>
:TRIGger[:SEQuence]:EXTernal[1]:DELay?
Set the amount of trigger delay when using the front panel external
trigger input. Set the trigger value to zero (0) seconds to turn off trigger
delay.
Factory Preset
and *RST: 0.0 s
Range: −500.0 ms to 500.0 ms
−100.0 ms to 500.0 ms for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit: seconds
Front Panel
Access:
Mode Setup, Trigger, Ext Front, Delay