206 Chapter5
Language Reference
CALibration Subsystem
External Signal Power for Internal 50 MHz Amplitude
Reference Alignment
:CALibration:REF50:AMPL <power>
:CALibration:REF50:AMPL?
You must set this value equal to the actual amplitude of the external 50
MHz amplitude reference signal applied to the RF INPUT connector.
This is used for aligning the 50 MHz amplitude reference with
CAL:REF50.
Preset
and *RST: −25.00 dBm
Range: −30 to −20 dBm
Default Unit: dBm
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command.
Front Panel
Access:
System, Alignments, Align subsystem, Align 50 MHz
Reference
Internal 50 MHz Amplitude Reference Alignment Control
:CALibration:REF50:ANOW
Immediately does the automatic alignment of the internal 50 MHz
amplitude reference oscillator. This command is used with the
interactive mode of the 50 MHz alignment, i.e. CAL:REF50:ENTer.
Remarks: You must be in the Service mode to use this command.
Use INSTrument:SELect.
A valid service password needs to be entered prior to
sending this command.
Front Panel
Access: System, Alignments, Align subsystem, Align 50 MHz
Reference