204 Chapter5
Language Reference
CALibration Subsystem
Load the Factory Default Calibration Constants
:CALibration:LOAD:DEFault
Load the factory default alignment data, ignoring the effect of any
alignments already done.
Front Panel
Access: System, Alignments, Restore Align Defaults
Align the Wide LC Prefilter
:CALibration:PFILter:LCWide
:CALibration:PFILter:LCWide?
Align the wide LC prefilter. (1.2 MHz to 7.5 MHz)
Remarks: A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Diagnostics
Align the Narrow LC Prefilter
:CALibration:PFILter:LCNarrow
:CALibration:PFILter:LCNarrow?
Align the narrow LC prefilter. (200 kHz to 1.2 MHz)
Remarks: A valid service password needs to be entered prior to
sending the command.
Front Panel
Access:
System, Alignments, Align Subsystem, IF
Align the Wide Crystal Prefilter
:CALibration:PFILter:XTALWide
:CALibration:PFILter:XTALWide?
Align the wide crystal prefilter. (20 kHz to 200 kHz)
Remarks: A valid service password needs to be entered prior to
sending the command.
Front Panel
Access: Enter service password and press
System, Diagnostics