Chapter 5 343
Language Reference
SENSe Subsystem
Waveform—Averaging Type
[:SENSe]:WAVeform:AVERage:TYPE
LOG|MAXimum|MINimum|RMS|SCALar
[:SENSe]:WAVeform:AVERage:TYPE?
Select the type of averaging.
Log - The log of the power is averaged. (This is also known as video
averaging.)
Maximum - The maximum values are retained.
Minimum - The minimum values are retained.
RMS - The power is averaged, providing the rms of the voltage.
Factory Preset
and *RST: RMS
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Resolution BW
[:SENSe]:WAVeform:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:WAVeform:BANDwidth|BWIDth[:RESolution]?
Set the resolution bandwidth. This value is ignored if the function is
auto-coupled.
Factory Preset
and *RST: 100.0 kHz for NADC, PDC, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), basic, service
500.0 kHz for GSM
2.0 MHz for cdmaOne
Range: 1.0 kHz to 5.0 MHz
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals