Chapter 5 343
Language Reference
SENSe Subsystem
Waveform—Averaging Type
[:SENSe]:WAVeform:AVERage:TYPE
LOG|MAXimum|MINimum|RMS|SCALar
[:SENSe]:WAVeform:AVERage:TYPE?
Select the type of averaging.
Log - The log of the power is averaged. (This is also known as video
averaging.)
Maximum - The maximum values are retained.
Minimum - The minimum values are retained.
RMS - The power is averaged, providing the rms of the voltage.
Factory Preset
and *RST: RMS
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Resolution BW
[:SENSe]:WAVeform:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:WAVeform:BANDwidth|BWIDth[:RESolution]?
Set the resolution bandwidth. This value is ignored if the function is
auto-coupled.
Factory Preset
and *RST: 100.0 kHz for NADC, PDC, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), basic, service
500.0 kHz for GSM
2.0 MHz for cdmaOne
Range: 1.0 kHz to 5.0 MHz
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.