296 Chapter5
Language Reference
SENSe Subsystem
Time Slot number
[:SENSe]:CHANnel:SLOT <integer>
[:SENSe]:CHANnel:SLOT?
Select the slot number that you want to measure.
In GSM mode the measurement frame is divided into the eight
expected measurement timeslots. Optimum alignment of these
measurement timeslots with the actual data timeslots may require
some trigger time delay. A trigger delay of about 20 ms is a reasonable
offset to use for a typical signal.
Factory Preset
and *RST: 0 for GSM, PDC mode
1 for NADC mode
Range: 0 to 5 for PDC mode
1 to 6 for NADC mode
0 to 7 for GSM mode
Remarks:
You must be in EDGE(w/GSM), GSM, NADC, PDC
mode to use this command. Use INSTrument:SELect to
set the mode.
Front Panel
Access:
Mode Setup, Radio, Frequency Hopping Repetition Factor