298 Chapter5
Language Reference
SENSe Subsystem
Training Sequence Code (TSC) Auto
[:SENSe]:CHANnel:TSCode:AUTO OFF|ON|0|1
[:SENSe]:CHANnel:TSCode:AUTO?
Select auto or manual control for training sequence code (TSC) search.
With auto on, the measurement is made on the first burst found to have
one of the valid TSCs in the range 0 to 7 (i.e. normal bursts only). With
auto off, the measurement is made on the 1st burst found to have the
selected TSC.
Factory Preset
and *RST: Auto
Remarks: Global to the current mode.
You must be in the EDGE(w/GSM), GSM mode to use
this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
FREQUENCY Channel, TSC (Std)