EasyManua.ls Logo

Agilent Technologies E4406A VSA Series - Sense]:Chpower:frequency:span Freq

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 301
Language Reference
SENSe Subsystem
Channel Power—Span
[:SENSe]:CHPower:FREQuency:SPAN <freq>
[:SENSe]:CHPower:FREQuency:SPAN?
Set the frequency span that will be used.
Factory Preset
and *RST: 2.0 MHz for Basic, cdmaOne, cdma2000
6.0 MHz for W-CDMA (3GPP), W-CDMA (Trial & Arib)
Range: 1.0 kHz to 10.0 MHz
Default Unit: Hz
Remarks: You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or Basic mode to use
this command. Use INSTrument:SELect to set the
mode.
Channel Power—Data Points
[:SENSe]:CHPower:POINts <integer>
[:SENSe]:CHPower:POINts?
Set the number of data points that will be used. Changing this will
change the time record length and resolution BW that are used.
Factory Preset
and *RST: 512
Range: 64 to 32768, in a 2
n
sequence
Remarks: You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or Basic mode to use
this command. Use INSTrument:SELect to set the
mode.