EasyManuals Logo

Agilent Technologies E4406A VSA Series User Manual

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
184 Chapter5
Language Reference
CALCulate Subsystem
Example: Select the spectrum measurement.
Use CALC:DATA4:PEAK? -40,10,FREQ to identify the
peaks above -40 dBm, with excursions of at least 10 dB,
in order of increasing frequency.
Query Results: Returns a list of floating-point numbers. The first value
in the list is the number of peak points that follow. A
peak point consists of two values: a peak amplitude
followed by the its corresponding frequency (or time).
If no peaks are found the peak list will consist of only
the number of peaks, (0).
The peak list is limited to 100 peaks. Peaks in excess of
100 are ignored.
Remarks: This command uses the data setting specified by the
FORMat:DATA command and can return real 32-bit,
real 64-bit, or ASCII data. The default data format is
ASCII.
History: Added in revision A.03.00 and later

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals