282 Chapter5
Language Reference
SENSe Subsystem
Adjacent Channel Power—Sweep Time
[:SENSe]:ACP:OFFSet:LIST:SWEep:TIME
<seconds>,<seconds>,<seconds>,<seconds>,<seconds>
[:SENSe]:ACP:OFFSet:LIST:SWEep:TIME?
Selects a specific sweep time. If you increase the sweep time, you
increase the length of the time data captured and the number of points
measured. You might need to specify a specific sweep speed to
accommodate a specific condition in your transmitter. For example, you
may have a burst signal and need to measure an exact portion of the
burst.
Selecting a specific sweep time may result in a long measurement time
since the resulting number of data points my not be the optimum 2
n
.
Use [:SENSe]:ACP:SWEep:TIME to set the number of points used for
measuring the reference channel.
You can turn off (not use) specific offsets with the
SENS:ACP:OFFSet:LIST:STATe command.
Factory Preset
and *RST:
Range: 1 µs to 50 ms
Default Unit: seconds
Remarks: You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History: Revision A.03.00 or later, in cdmaOne revision A.04.00
Mode Offset A Offset B Offset C Offset D Offset E
Basic &
cdmaOne
11.20 ms 11.20 ms 11.20 ms 11.20 ms 11.20 ms

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals