Chapter 5 341
Language Reference
SENSe Subsystem
Waveform—ADC Range
[:SENSe]:WAVeform:ADC:RANGe
AUTO|APEak|APLock|GROund|M6|P0|P6|P12|P18|P24|
[:SENSe]:WAVeform:ADC:RANGe?
Select the range for the gain-ranging that is done in front of the ADC.
This is an Advanced control that normally does not need to be changed.
Auto - automatic range
Auto Peak - automatically peak the range
Auto Peak Lock - automatically peak lock the range
Ground - ground
M6 - subtracts 6 dB of fixed gain across the range
P0 to 24 - adds 0 to 24 dB of fixed gain across the range
Factory Preset
and *RST: Auto
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Number of Averages
[:SENSe]:WAVeform:AVERage:COUNt <integer>
[:SENSe]:WAVeform:AVERage:COUNt?
Set the number of sweeps that will be averaged. After the specified
number of sweeps (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST: 10
Range: 1 to 10,000
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.