EasyManuals Logo

Agilent Technologies E4406A VSA Series User Manual

Agilent Technologies E4406A VSA Series
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #341 background imageLoading...
Page #341 background image
Chapter 5 341
Language Reference
SENSe Subsystem
Waveform—ADC Range
[:SENSe]:WAVeform:ADC:RANGe
AUTO|APEak|APLock|GROund|M6|P0|P6|P12|P18|P24|
[:SENSe]:WAVeform:ADC:RANGe?
Select the range for the gain-ranging that is done in front of the ADC.
This is an Advanced control that normally does not need to be changed.
Auto - automatic range
Auto Peak - automatically peak the range
Auto Peak Lock - automatically peak lock the range
Ground - ground
M6 - subtracts 6 dB of fixed gain across the range
P0 to 24 - adds 0 to 24 dB of fixed gain across the range
Factory Preset
and *RST: Auto
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Number of Averages
[:SENSe]:WAVeform:AVERage:COUNt <integer>
[:SENSe]:WAVeform:AVERage:COUNt?
Set the number of sweeps that will be averaged. After the specified
number of sweeps (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST: 10
Range: 1 to 10,000
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4406A VSA Series and is the answer not in the manual?

Agilent Technologies E4406A VSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4406A VSA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals