Anatel A643a - Troubleshooting 177 of 220
RPS - June 2007 - Edition 12
Anatel Operator Manual
Analysis Alarm Codes—
Code Instrument Display Indication
21 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
22 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
23 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
24 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
25 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
26 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
27 Conductivity Range Water has drained from the analysis cell or the current sample vial is dry. Ensure that water
is flowing normally through the instrument and it is in the correct orientation. If the issue
persists, call the
Anatel Service Department.
28 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
29 Water Conditions Water has drained from the analysis cell or the current sample vial is dry. Ensure that water
is flowing normally through the instrument and it is in the correct orientation. If the issue
persists, call the
Anatel Service Department. TOC detection limits may be too low for
current water conditions. If the issue persists, call the
Anatel Service Department.
30 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
31 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
32 See Guide
Possible hardware performance issue. Call the
Anatel Service Department.
33 Check UV Lamp Replace the UV lamp.
34 Water Conditions Possible hardware performance issue or TOC detection limits may be too high for current
water conditions.
35 Water Conditions Possible hardware performance issue or TOC detection limits may be too low for current
water conditions.
36 Water Conditions Current water conditions interfere with the TOC measurement.
37 Water Temp Sample water temperature < 5 °C or possible hardware performance issue. A heat
exchanger may be required. If issue persists, call the
Anatel Service Department.
38 Water Temp Sample water temperature >95 °C or possible hardware performance issue. A heat
exchanger may be required. If issue persists, call the
Anatel Service Department.
39 Conductivity Range Conductivity of the water sample is too high for TOC measurement. Water has drained from
the analysis cell or the current sample vial is dry. Ensure that water is flowing normally
through the instrument and it is in the correct orientation. If the issue persists, call the
Anatel Service Department.
40 Conductivity Range Conductivity of the water sample is too low for TOC measurement. Water has drained from
the analysis cell or the current sample vial is dry. Ensure that water is flowing normally
through the instrument and it is in the correct orientation. If the issue persists, call the
Anatel Service Department.
42 UV Lamp Replace the UV lamp.
Table 14-1 :
Anatel A643a Alarm Codes (Continued)