CAPINTEC, INC. CAPTUS
®
600
ENTERING CAPSULE DATA
The One Capsule Data Entry Screen (Figure 9-1) will appear.
LET
OK
KETCHUM, LIN
PROTOCOL: PROT 1
NUCLIDE: I 131
RX ID :>
NO. CAPS: 1 TO MEASURE
Figure 9-1
At the “RX ID” prompt, input the name or identification number of the capsule to be
administered to this patient. The “RX ID” can consist of up to 19 numbers and/or letters.
Press F1 [NUM] to switch to numeric input and display “LET” for F1.
Press F1 [LET] to switch back to letter input and display “NUM” for F1.
The default of 1 capsule to be measured is displayed.
The number of capsules to measure depends on the Multi-Capsule option:
“NO” or “ONE”: The number of capsules to be measured is always 1 and cannot be
changed.
“EACH” or “ALL”: The number of capsules can be changed.
Press the F4 [OK] key when the data is correct.
Figure 9-2 will appear.
KETCHUM, LIN
PROTOCOL: PROT 1
NUCLIDE: I 131
RX ID : I1234
NO. CAPS: 1 TO MEASURE
DONE
MEAS
Figure 9-2
9-2 THYROID UPTAKE: ONE CAPSULE METHOD November 07