CAPINTEC, INC. CAPTUS
®
600
CHAPTER 4
SYSTEM SETUP
INSTALLATION
Your system has been shipped from the factory pre-configured and fully tested. Installation
consists of placing the components on the stand and securing them in place as shown in the
CAPTUS
®
600 Unpacking and Assembly Instructions document.
ENVIRONMENT REQUIREMENTS
Operational
The instrument should be located where the level of the background radiation is as low and
as constant as possible.
The instrument should be located where the temperature is stable within a range of +50°F to
+85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-condensing to
warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of 27–31
inches of mercury (91–105 kilopascals).
Storage
The instrument should be stored where the temperature is stable and the range is from
+39°F to +110°F (+4°C to +43°C) and the maximum relative humidity is 90% non-condensing
to warrant maximum reliability.
The instrument should be stored where the barometric pressure is within a range of 15–33
inches of mercury (51–112 kilopascals).
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
November 07 SYSTEM SETUP 4-1