CAPINTEC, INC. CAPTUS
®
600
Cs137 STANDARD SETUP
All calibration of the Capintec CAPTUS
®
600 is done with a Cesium-137 standard reference
source. Before using this source to calibrate the system, the source calibration data must be
input in the CAPTUS
®
600 memory.
Press the HOME key then the UTIL key. The main Utilities menu (
Figure 5-10) will appear.
UTILITIES
1. MULTICHANNEL ANALYZER
2. STAFF BIO ASSAY RESULTS
3. SETUP
4. DIAGNOSTICS
0. RETURN TO PROBE SCREEN
14:32
Figure 5-10
Press 3 to select SETUP. The message “PASSWORD:” will appear. Type the last three digits
of your CAPTUS
®
600 serial number, and press the ENTER key. The Setup menu (Figure 5-
11
) will appear.
1. Ci/Bq
2. TEST DAT
A
3. THYROID UPTAKE
4. BIO ASSAY
6. ACTIVITY CONVERSION FACTORS
SETUP
5. AUTO CALIBRATION
7. WIPE AND MCA SETUP
8. OTHE
R
0. RETURN TO UTILITIES MENU
14:32
Figure 5-11
Press 2 to select TEST DATA. The Test Data menu (
Figure 5-12) will appear.
November 07 SYSTEM INITIALIZATION 5-7