CAPINTEC, INC. CAPTUS
®
600
Collimator Block Inspection
Once a year, the Collimator Mounting Block (1) should be inspected for the following:
excessive play between the Spring-Arm shaft (2) and the Collimator Mounting Block,
excessive play between the Collimator pin (3) and the Collimator Mounting Block.
If there is excessive play or the knob(s) cannot be tightened, contact Capintec’s Authorized
Service Center for a information on replacement items, instructions or on-site service.
Figure 20-1
WARNING
FAILURE TO FOLLOW THESE INSTRUCTIONS COULD RESULT IN SERIOUS INJURY
OR PRODUCT MALFUNCTION.
SERVICING
The system is covered by a one year limited warranty, under normal conditions of use.
Other than the Readout Unit and Isolation Transformer’s fuses (see Fuse Servicing in this
chapter), there are no user serviceable parts contained in the system.
Every five years, the system should be returned to Capintec’s Authorized Service Center for
a complete verification.
CAPINTEC, Inc.
620 Alpha Drive
Pittsburgh, PA 15238
Phone (412) 963-1988, 1-800-227-6832
Fax (412) 963-0610
November 07 CLEANING AND MAINTENANCE 20-3