CAPINTEC, INC. CAPTUS
®
600
If the prompt is not at the first uptake, the previous uptake can be re-measured.
If the “ONE CAPSULE” method has been chosen and the capsule has been measured but
not administered, it can be re-measured.
Use the ARROW keys to move the prompt to the correct line if necessary.
Press ENTER to begin the measurement.
If all the scheduled measurements have been made,
Figure 8-12 will appear.
14:32
THYROID UPTAKE
DASILVA, AND
PROTOCOL: 1 CAPSULE
CAPSULE : TU 11:09 250.0
UPTAKE 1: 24%
UPTAKE 2: 30%
> REVIEW RESULTS
MEASUREMENT FINISHED
Ci
Figure 8-12
Pressing ENTER with the prompt at “REVIEW RESULTS” will bring up the Graphic Display
Screen (
Figure 8-14) described below.
MEASURING THE NECK AND THIGH
The main part of the Thyroid Uptake Study is measuring the patient’s neck and thigh. This is
the same for all Calculation Methods.
The measurement begins by prompting the user to measure the neck at the distance
previously chosen as shown in
Figure 8-13.
November 07 THYROID UPTAKE GENERAL INSTRUCTIONS 8-7