CAPINTEC, INC. CAPTUS
®
600
DISPLAY CONTRAST ADJUSTMENT
This adjustment controls the contrast of the Readout LCD. Perform adjustment until LCD
screen displays the best contrast.
DATA ENTRY INPUT LIMITS
Thyroid Uptake Protocols
REGIONS OF INTEREST
I123 I131 Tc99
Left of peak 10 – 40% 10 – 60% 10 – 40%
Right of peak 10 – 25% 10 – 25% 10 – 25%
COUNTING TIME 10-999 seconds
NUMBER OF TESTS 1 – 4
NUMBER OF REPETITIONS 1 or 2
Thyroid Uptake Measurements
DISTANCE 15 – 40 cm
NUMBER OF CAPSULES 1 – 4
DOSE & THYROID UPTAKE STANDARD
I123
10. – 600. μCi
0.37 – 22.2 MBq
I131
1. – 9973. μCi
0.037 – 369. MBq
Tc99m
1. – 9973. μCi
0.037 – 369. MBq
DOSE
Rx ID NUMBER 19 characters
THYROID UPTAKE STANDARD
ID NUMBER 10 characters
Cs137 Standard
ACTIVITY LIMITS
PROBE
.1 - 250. μCi
3.7 - 9250. kBq
WELL
.01 - 1.0 μCi
0.37 - 37. kBq
Note: The source must be positioned so that the maximum count rate is not
exceeded.
November 07 GENERAL OPERATING OVERVIEW 3-11