CAPINTEC, INC. CAPTUS
®
600
CHAPTER 13
WIPE TEST
INTRODUCTION
Before performing Wipe Tests, the Calibration Test must have been performed for the
calendar day. Reference CHAPTER 6: ACCEPTANCE & QUALITY ASSURANCE TESTS for
instructions on performing the Calibration Test.
The room background must be measured in the Well for the 400 keV and 1600 keV scales
once a day.
If you are performing a Well Function (e.g. a lab test), press HOME.
Figure 13-1 will appear.
If you are performing a Probe Function (e.g. Thyroid Uptake), press HOME and then press
WELL.
Figure 13-1 will appear.
13:06
WELL
WIPE
F1 F2 F3 F4
LAB
Figure 13-1
If your system is configured for just one detector and you press the WELL key, the system
will automatically use the detector connected to the PROBE INPUT on the back panel.
Press the F1 [WIPE] key.
MEASURING BACKGROUND
If the Calibration Test has not been performed in the current calendar day, a message
instructing the user to perform Test will appear on the display.
Figure 13-2 will appear
instructing you to perform background. If background has already been measured in the
current calendar day, the Wipe Test Area Select Screen (
Figure 13-5) will appear.
November 07 WIPE TEST 13-1