EasyManua.ls Logo

Capintec CAPTUS 600 - Activity of a Nuclide; Exiting MCA Measurement

Capintec CAPTUS 600
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
600
Activity of a Nuclide
The feature is only available if Calibration Test has been performed during the current
calendar day.
The counts in a Region of Interest corresponding to a given nuclide can be obtained by
pressing one of the Pre-Set Nuclide keys (
Cr51, Co57, Ga67, Tc99m, In111, I131, I123,
I125, Tl201), a USER key which has had a nuclide assigned to it or the NUCL key. After the
NUCL key is pressed, enter the name of the nuclide and press ENTER.
The counts in the ROI are displayed on the top line.
Exiting MCA Measurement
To exit the MCA Results Screen, press the BSP key, and the display will return to the
Parameter Entry Screen.
19-8 MULTICHANNEL ANALYZER OPERATION November 07

Table of Contents

Related product manuals