CAPINTEC, INC. CAPTUS
®
600
For example, to enter Cobalt-57, input C, O, 5, 7 and ENTER; to enter Iodine-131, input I,
SP, 1, 3, 1 and ENTER.
Press F4 [OK] when the Nuclide Name is correct.
Press F1 [DONE] if all CFs have been input or measured.
Figure 5-23 will appear.
Note: If a CF for the chosen nuclide already exists, the current value will be displayed and
you will be asked if you would like to change it. If you respond that you would like to
change it, the system will proceed to
Figure 5-23, otherwise you will return to Figure
5-22
.
Selecting Measure or Input CF
13:06
CF FOR WELL
NUCLIDE NAME: TC99M
MEASURE CF? (Y OR N)
Figure 5-23
Press Y (YES) to measure the CF.
Press N (NO) to input a CF.
Input CF
If INPUT CF was selected, the Input CF screen (
Figure 5-24) will appear.
5-14 SYSTEM INITIALIZATION November 07