CAPINTEC, INC. CAPTUS
®
600
This question, with the appropriate Standard Name in the second line, will appear on the
screen:
DO YOU WANT TO DELETE
STANDARD NAME
Y OR N
If a TU Standard is being used in a TU study, then it cannot be deleted. If an attempt is made
to delete a Standard that is in use, the following message is displayed:
STANDARD IS BEING USED
CANNOT DELETE
Before deleting the Standard, the patients’ studies must be completed and the patients
deleted.
PRINTING TU STANDARD DATA
Press PRINT [P] on the Standard List Screen to obtain a printout of the TU Standard data
that has been entered.
MEASURING TU STANDARD
The TU Standard must be measured once a day (before it can be used in a patient study).
When the Standard List Screen (
Figure 11-3) appears, move the prompt to the TU Standard
to be measured and press F4 [MEAS]. The Standard Measurement Screen (
Figure 11-6) will
appear.
TU STANDARD
START
14:32
MEASURE I 131
STANDARD
IN
ID: EE
AT:>25.0 CM
Counting Repetition #1 / 2
Figure 11-6
The default distance of 25 cm is displayed.
11-4 THYROID UPTAKE: STANDARD CAPSULE METHOD November 07