EasyManua.ls Logo

Capintec CAPTUS 600 - Page 125

Capintec CAPTUS 600
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
600
Input the distance that will be used for the TU studies. The distance entered here cannot be
changed when a patient is being measured.
When the distance is correct, place the capsule in the phantom and press F4 [START]. The
Standard Count Screen (
Figure 11-7) will appear.
200
19.56kcpm
A
BORT
OK?
(Y OR N)
400 keV
STANDARD
Figure 11-7
If the measurement is acceptable, press Y and ENTER. Otherwise, press N and ENTER and
the unit will count again.
The unit will then prompt you to measure room background. Remove all sources and press
the F4 [START] key. Accept or reject the room background with a Y or N.
The Standard Measurement Results Screen (
Figure 11-8) will appear.
14:32
TU STANDARD
STANDARD : 1504 cpm
CONV FACTOR : 15.04 cpm/
OK? (Y OR N)
μ
Ci
Figure 11-8
Respond Y for “YES” and N for “NO”.
If the measurement is not accepted, the TU Standard Measurement process will begin again
from Standard Measurement Screen (
Figure 11-6).
November 07 THYROID UPTAKE: STANDARD CAPSULE METHOD 11-5

Table of Contents

Related product manuals