CAPINTEC, INC. CAPTUS
®
600
• DETECTORS: Use a soft cloth that is lightly dampened with water, window cleaner,
or a mild detergent. Wipe the front surface and or the well as necessary to remove
smudges of surface contamination.
Note: If contamination is suspected, dispose of the cleaning cloth in an approved
container for contaminated materials.
• POINTER ASSEMBLY: Use a soft cloth that is dampened with propanol alcohol to
sterilize or clean the black pointer and assembly.
• READOUT: Wipe down with a damp cloth; do not use solvents or aerosol cleaners.
• PRINTER: Wipe down with a damp cloth; do not use solvents or aerosol cleaners.
Wipe the outside surfaces of the printer case to remove the dust, smudges, and
stains. The interior of the printer does not require cleaning.
PREVENTATIVE MAINTENANCE
The Quality Assurance Tests described in CHAPTER 6: ACCEPTANCE & QUALITY
ASSURANCE TESTS, SECTION: QUALITY ASSURANCE TESTS, should be performed
periodically as stated.
Tests must be performed in an environment where the temperature is stable within a range
of +50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing. The unit should be powered-up for at least one-half hour prior to performing any
measurements. No other precautions need to be observed.
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
Note: Never put a sharp instrument like a pen into the Well or Collimator. If the aluminum
coating of the NaI Detector is scratched, allowing light and moisture to penetrate, the
detector will be ruined. The rod source is plastic and can be safely placed against the
surface of the detectors when handled carefully.
If the unit fails to pass Auto Calibration Test or any other test, the user
should not attempt to
perform any adjustments to the system. In this event, please contact Capintec’s Authorized
Service Center for further assistance.
The Quality Assurance tests should be immediately performed if:
• The equipment has been subjected to extreme physical stress,
• Liquids enter the readout unit, and/or Probe and Well detectors, or
• Any cable shows signed of damage
20-2 CLEANING AND MAINTENANCE November 07