CAPINTEC, INC. CAPTUS
®
600
VIEWING RESULTS
After the neck and thigh have been measured, the Results Screen (Figure 9-13) will appear.
CAPSULE (C) : 2915 cpm
NECK (N) : 2400 cpm
THYROID (T:N-B) : 1207 cpm
DECAY FACTOR (D) : 0.993
NUMBER CAPS (K) : 1
ENTER INITIALS:
UPTAKE (T/KCD) : 41.7%
THIGH (B) : 1193 cpm
Figure 9-13
Input your initials and press ENTER. The Graphics Screen (
Figure 9-14) will appear.
AGAIN
DONE
P
............................................
............................................
............................................
............................................
...............
.
............................
2
24 hours
50
40
30
20
10
*
%
U
P
T
A
K
E
DASILVA, AND
*
Figure 9-14
Press PRINT [P] to print a report of this Thyroid Uptake study. Refer to
Figure 9-15 for an
example of the printout.
Press F1 [AGAIN] if an error was made in the measurement and the measurement needs to
be repeated.
Press F4 [DONE] to exit the screen and return to the Patient Selection Screen.
Note: Refer to CHAPTER 8: THYROID UPTAKE: GENERAL INSTRUCTIONS for a more
detailed discussion of this display.
November 07 THYROID UPTAKE: ONE CAPSULE METHOD 9-9