CAPINTEC, INC. CAPTUS
®
600
Isolation Transformer
An isolation transformer is provided with the CAPTUS
®
600 system to ensure that the
leakage current meets EN60601-1-1 Medical Equipment Safety Standard. A detachable
4.6m (15 ft) power cord, with a hospital grade plug, is provided to connect the isolation
transformer to the power source. Dual power line fuses are provided at the power entry
module for quick replacement.
Dimensions: Height 13.3 cm (5.25")
Width 17.5 cm (6.88")
Depth 10.9 cm (4.31")
Printer
Okidata 184 Turbo
80 column dot matrix printing mode.
Graphics and near letter quality text printing
Power Requirements: 115V 0.4A 50/60Hz
Dimensions: Height 8.1 cm (3.2")
Width 36.0 cm (14.2")
Depth 27.4 cm (10.8")
Weight 4.5 kg (10 lb.)
Readout Unit
256 channel MCA with LCD display
Power Requirements: 30 Watts
115V 0.3A 50/60Hz
Power Conditioning: Power Line Filter
Transient Voltage Suppressers
Dual Power Line Fuses
Dimensions: Height 16.5 cm (6.5")
Width 26.7 cm (10.5")
Depth 35.9 cm (14.125")
Weight 4.5 kg (10 lb.)
Nuclear Data
9 Pre-set nuclide data keys
5 User assignable nuclide data keys
Over 60 nuclides in memory (major gamma-ray energies and intensities, half-life)