CAPINTEC, INC. CAPTUS
®
600
Press the F3 [DONE] key if you are not ready to measure the capsule(s). The screen returns
to the TU Patient Selection Screen, with the new patient added to the list.
Press the F4 [MEAS] key to measure the capsule(s).
MEASURING THE CAPSULE(S)
The Capsule Measurement Screen (Figure 9-3) will appear with instructions for measuring
the capsule(s) in the phantom.
14:32
THYROID UPTAKE
START
KETCHUM, LIN
PROTOCOL: PROT 1
MEASURE I 131
1 CAPSULE
IN
AT: >25.0 CM
COUNTING REPETITION # 1 / 2
EXIT
Figure 9-3
The instructions for measuring the capsules depends on the Multi-Capsule option:
• “NO” or “ONE”: Measure 1 capsule in the phantom.
• “ALL”: Measure the number of capsules that was entered in the One
Capsule Data Entry Screen.
• “EACH”: Measure the #1 capsule, then the #2 capsule until the number of
capsules that was entered in the One Capsule Data Entry Screen
has been measured. Each capsule is measured for the number of
repetitions in the protocol.
If Pre-Dose Measurement has been performed, the distance displayed will be the distance
entered for the Pre-Dose Measurement and the prompt (>) will not be displayed.
If Pre-Dose Measurement has not been performed, the displayed “AT:” distance of 25.0 cm
is the default value of the Probe’s position away from the patient's neck. If this is not the
value to be used for the Thyroid Uptake study, input the correct value. This value will be used
throughout the study and cannot be changed.
Place the capsule(s) in the phantom and press F4 [START] to begin the measurement.
The Counting Screen (
Figure 9-4) will appear.
November 07 THYROID UPTAKE: ONE CAPSULE METHOD 9-3