CAPINTEC, INC. CAPTUS
®
600
13:06
CF FOR WELL
OK
NUCLIDE NAME: TC99m
INPUT CF: >
Figure 5-24
Input the CF in units of dpm/cpm.
Press F4 [OK] when the value is correct. The Nuclide Name screen (
Figure 5-23) will
re-appear.
Note: This must be a value that had been previously measured in the CAPTUS
®
600
using the same detector and source configuration.
Measure CF
If MEASURE CF was selected, the CF Source Calibration Data screen (
Figure 5-25)
will appear:
13:06
CF FOR WELL
OK
NUCLIDE NAME: TC99m
CALIB ACTIVITY: >
CALIB DATE :
CALIB TIME : 12:00
Ci/Bq
Ci
Figure 5-25
Input the calibration activity, date and time. If it is a long-lived source, the time will not
be known and can be left as 12:00.
Press F1 [Ci/Bq] to switch between entering the activity in Ci to Bq and back. This
will not affect the unit used for anything else.
Press F4 [OK] when all entries are correct.
Figure 5-26 will appear.
November 07 SYSTEM INITIALIZATION 5-15