CAPINTEC, INC. CAPTUS
®
600
Conversion Factors
Conversion Factors that have been input or measured for each nuclide. Conversion
Factors may be different for Well or a Probe. Both Conversion Factors are printed.
Additional Nuclides
A listing of nuclide data for any additional nuclides that have been entered.
User Keys
A listing of nuclides assigned to the USER keys.
Standard Source
A listing of calibration date, calibration activity and current activity of the Standard
Cs137 sources that have been entered in the CAPTUS
®
600 memory.
HV Reference (Probe and Well)
A listing of the High Voltage Reference set during the Autocalibration Test. These
values will vary between 0 and 255.
Fine Gain Adjustment (Probe and Well)
A listing of the Fine Adjustment set during Calibration Test. These values will vary
between 0 and 255.
PROM Values
These diagnostic values uniquely determine the software revision level. If the system
reads these values incorrectly, the PROM test will fail.
Memory Tests
The results of the RAM, PROM, and EEPROM tests. If the PROM test fails, the
PROM values will be printed. Any test that fails indicates a system malfunction and
warrants a call to Capintec’s Authorized Service Center.
6-14 ACCEPTANCE & QUALITY ASSURANCE TESTS November 07