ECG
1.SelecttheHRparameterwindow>Setup>Page1.
2.SelectDeactivateECGLeadsOff.
Settingthebeatvolume
1.SelecttheHRparameterwindow>Setup>Page2.
2.SetthebeattonevolumewiththeBeatVolumeselector.
Therangeis0(volumeoff)to10.
Selectingthebeatsource
1.SelecttheHRparameterwindow>Setup>Page2.
2.SelectthebeatsourcefromtheBeatSourcelist:
●PrimaryHR(ThebeatsourceisthesameastheprimaryHRsourceselection.)
●ECG(ThebeatsourceisECGregardlessoftheprimaryHRsourceselection.)
●Invasivepressurechannel(s)labeledArtX,FemX,orUACX,whereX=channel
number.
●Pleth(SpO2)
Thebeatsourceindicator
willappearbesidethechosenbeatsourceonthescreen,
andthebeatsoundwillreectthebeatofthatsource.
Settingthebeeptoneduringbradycardiaalarms
NOTE
NICUsoftwarepackageonly.
ThisselectionisavailableiftheBeatVolumeforQRSissetto0(off).
Ifthealarmforbradycardiahasbeensettoofforsilenced,ortheECGalarmsare
silencedpermanently,thentheQRStonedoesnotsound,either.
1.SelecttheHRparameterwindow>Setup>Page2.
2.SelecttheBeatToneonBradyOnlycheckboxtoenablethebeeptone.
Whenthebeeptoneison,theQRStoneswillsoundonlywithBradyalarm
conditions.
●Ifthealarmvolumehasbeensettobelow8,theQRStoneswillsoundatthe
selectedalarmvolumelevel+2morelevels.
●Ifthealarmvolumehasbeensetto8ormore,theQRStoneswillsoundat
alarmvolumelevel10.
SelectingtheECGwaveformsize
ThisselectionadjuststhesizeofthedisplayedECGwaveform.
1.SelecttheHRparameterwindow>Setup>Page2.
2.SelectavaluefromtheSizelist.
Theselectionsare0.5x,1x,2x,4x.Thesmallerthevalue,thesmallerthewaveform.